Philips HP 50XM M1360b Fetal Monitor

The HP/Philips Fetal Monitor Series 50XM is a high-quality unit utilized during delivery and labor. It provides continuous maternal vital signs monitoring and maternal non-invasive blood pressure readings. It features individual twin’s heart rate monitoring as well as uterine activity calculated in relation to a selected baseline. The 50XM
provides signal quality indicators for each fetal heart channel for optimal transducer placement. The color-coded sockets match the transducer cables for easy identification.

  • Continuous maternal vital signs monitoring, and maternal non-invasive blood pressure
  • Individual twins heart rate monitoring
  • Uterine activity calculated in relation to a selected baseline
  • Signal quality indicators for each fetal heart rate channel for optimal transducer placement
  • Toco and intrauterine pressure monitoring
  • Color-coded sockets match transducer cables.

Philips HP 50 XM Specifications

Dimensions

  • Height: 5.8 in (147mm)
  • Width: 16.6 in (422 mm)
  • Depth: 15.4 in (392mm)
  • Weight: 31.96 lbs (14.6 kg)

Environmental

  • Operating Temperature: 0°C to + 55°C
  • Storage Temperature: -40°C to +75°C
  • Relative Humidity: 5% to 95%

Fetal Display

  • Heart Rate Range
    • US: 50 to 240 bpm
    • DEGG: 30 to 240 bpm
    • MHR: 30 to 240bpm (not displayed)
  • External Toco Range: 0 to + 127 relative units
  • IUP Range: -99 to 127 mmHg
  • Fetal SpO2 Range: 0-99%
  • Fetal Heart Rate Alarm Limits
    • Bradycardia Alert Range: 60 to 120 bpm Adjustable in 10 bpm steps (default: 110 bpm)
    • Tachycardia Alert Range: 150 to 210 bpm adjustable in 10 bpm stems (Default: 150 bpm)
  • Fetal Heart Rate alarm Delay
    • Bradycardia Alert Delay: 10 to 300 sec adjustable in 10 sec steps (Default: 60 sec)
    • Tachycardia Alert Delay: 10 to 300 sec adjustable in 10 sec steps (Default: 60 sec)

Maternal non invasive Blood Pressure

  • Pressure Transducer Accuracy:
    • 15°C to 25°C ±3 mmHg
    • 10°C to 35°C ±3 mmHg (±0.6% of reading)
    • 0°C to 55°C ±3 mmHg (±1.7% of reading
  • Measurement Range:
    • Systolic: 30 to 270 mmHg
    • Diastolic: 10 to 245 mmHg
  • Cuff inflation rate: typically less than 10 seconds
  • Auto Mode Repetition time: 2, 5, 10, 15, 30, 60 minutes
  • Start Mode Duration: 5 minutes
  • Pulse Rate Range: Measurable within heart rate range of 30 to 240 bpm, averaged during NIBP measurement.

Power Requirements

  • Operating Voltage
    • 100 – 120 V (± 10%)
    • 220 – 240 V (±10%)
  • Line Frequency 50 to 60 Hz
  • Power Consumption 60 VA max

Fetal Display

  • Heart Rate Range
    • US 50 to 240 bpm
    • DECG 30 to 240 bpm
    • MHR 30 to 240 bpm (not displayed)
  • External Toco Range 0 to +127 relative units
  • IUP Range -99 to +127 mmHg
  • Fetal SpO2 Range 0 – 99%

Maternal Pulse Oximetry

  • Percentage Range 0 to 100%
  • BPM Range 30 to 300 bpm
  • Accuracy ± 1%
  • Resolution 1 bpm
  • Pulse Rate Limit Alarms
    • Range 70% – 90%
    • Adjustment 1% steps
  • Accuracy at 1 standard deviation
    • M1191A 70 to 99% ± 2.5%
    • M1192A 70 to 99% ± 2.5%
    • M1194A 70 to 99% ± 4%
    • Nellcor sensors 80 to 99% ± 3%

Maternal ECG and Heart Rate

  • Heart Rate Measurement
    • Range 30 to 240 bpm
    • Accuracy ±1 bpm
    • Display Resolution: 1bpm
    • Recorder Resolution: 0.25bpm
  • Heart Rate Alarm Limits (excluding NIBP)
    • Range 30 to 250 bpm
    • Adjustment 5 bpm steps